TandD 4-channel logger for thermocouples (types K, J, T, S, R, N, E). Measuring range -200 to +1760 °C depending on probe type. Modular — can be combined to 16 channels
4.362 SEK
TandD 4-channel logger for thermocouples (types K, J, T, S, R, N, E). Measuring range -200 to +1760 °C depending on probe type. Modular — can be combined to 16 channels
Volume discount: 5 pcs: 2% · 10 pcs: 3% · 25 pcs: 5%
- Free shipping on orders over 3000 SEK excl. VAT
- Data loggers with WiFi, USB and cloud service
- Call us +46 8 753 00 04
Other models in the series 16
Description
The 56-MCR-4TC is TandD’s 4-channel industrial thermocouple logger featuring a built-in touch panel, SD card storage, and galvanically isolated inputs. For larger installations, up to 4 units can be linked (16 channels total), which are synchronized via a common start and share settings.
Supports thermocouples K, J, T, S, R — covering temperatures from -199 to +1760 °C. The galvanic isolation between channels allows you to measure signals with different potential levels without crosstalk. Massive memory: up to 480,000 data points at 1 channel or 120,000 at 4 channels — this is a logger designed for long industrial processes.
NOTE: Thermocouples are NOT included — connect your own. Also available: 56-MCR-4V for voltage measurement (±50 V).
Thermocouple Compatibility
- Type K — -199 to +1370 °C (industrial standard)
- Type J — -199 to +1200 °C
- Type T — -199 to +400 °C (cryogenic)
- Type S — -50 to +1760 °C (PtRh10-Pt)
- Type R — -50 to +1760 °C (PtRh13-Pt)
Features
- 4 channels thermocouple with galvanic isolation
- Measurement range -199 to +1760 °C (sensor-dependent)
- Accuracy K/J/T: ±(0.5 + 0.003 × t) °C at ≥-100 °C; S/R: ±(1.5 + 0.003 × t) °C at ≥100 °C
- Cold Junction compensation ±0.5 °C (10–40 °C operating range)
- Resolution 0.1 °C
- Groupable — up to 4 units = 16 synchronized channels
- Massive memory: up to 480,000 / 240,000 / 160,000 / 120,000 measurements (1/2/3/4 channels)
- SD card storage (SD/SDHC) — automatic export when internal memory is full = no data loss
- Built-in touch panel for easy setup + trend graph display
- Preheating function saves battery life
- Logging interval from 100 ms to 60 min
- Powered by 2 × AA batteries or AC adapter or USB bus
Specification
| Measurement channels | Temperature 4 channels — thermocouple K/J/T/S/R via screw terminals, galvanically isolated |
|---|---|
| Measurement range | -199 to +1760 °C (sensor dependent) |
| Cold Junction compensation | ±0.5 °C (10–40 °C operation), ±0.8 °C (other) |
| Resolution | 0.1 °C |
| Input method | Scanning, differential input, channel isolated |
| Logging capacity | 480,000 (1 ch) / 240,000 (2 ch) / 160,000 (3 ch) / 120,000 (4 ch) measurements |
| Logging interval | 100, 200, 500 ms / 1, 2, 5, 10, 15, 20, 30 sec / 1, 2, 5, 10, 15, 20, 30, 60 min |
| Group function | Up to 4 units (16 channels) synchronized |
| Display | Built-in LCD with touch panel — displays measurements, trend graph, status |
| Communication | USB 2.0 Mini-B |
| External memory | SD / SDHC card (manual or automatic export) |
| Power supply | 2 × AA alkaline LR6, 2 × AA Ni-MH, AC adapter 56-AD-05A2/C2, USB bus 5V 250 mA |
| Battery life | approx. 4.5 to 130 days (4 channels, instant value, AA alkaline) |
Included
- 56-MCR-4TC logger with touch panel
- 2 × AA alkaline batteries
- USB Mini-B cable 56-US-15C
- Manual set with warranty certificate
- NOTE: thermocouples NOT included — ordered separately
- NOTE: SD card NOT included
Applications
The 56-MCR-4TC is selected for industrial temperature measurement using thermocouples: heat treatment furnaces, glass and metal industries, plastic recycling, heat-retaining processes, autoclaves, and calibration tasks requiring multiple thermocouples simultaneously. The group function (16 channels) makes it suitable for temperature mapping of large furnaces/chambers. For voltage measurement instead, see 56-MCR-4V. For a simpler 2-channel thermocouple model with WiFi, see 56-TR75A2.
Additional information
| Product Type | |
|---|---|
| Measures | Temperatuur |
| Communication | Bedraad USB |
| Tillbehör Loggers | 56-TR-75NW, 56-TR45, 56-TR75A2 |
| Tillbehör Annat | 56-AD-05C2, 56-AD-06C1, 56-SO-TD1 |