Hoppa till innehåll
Free shipping on orders over 3000 SEK excl. VAT! Volume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%
Volume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%The air in the fridge doesn’t show the temperature of the vaccineThe Air in the Fridge Lies: How to Monitor the Real Temperature of Your Goods Without False AlarmsHow to show in seconds whether the load stayed at the right temperature and humidity during transportThe air feels heavy long before anyone measures it: how to make carbon dioxide levels visibleThe freezer showed -18 °C on Monday morning, but what happened over the weekend?The core temperature is what counts, and it only shows if someone has logged itTwo measuring points and the difference between them: how to make the temperature differential in a process visibleFour Measuring Points and a Time-Stamped Log Show Where the Heat in the Process Actually GoesYou Can’t Guess the Core Temperature of a Product: How to Document It Without Standing Next to ItThe fridge looked like it was holding the right temperature, but what happened overnight?Volume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%The air in the fridge doesn’t show the temperature of the vaccineThe Air in the Fridge Lies: How to Monitor the Real Temperature of Your Goods Without False AlarmsHow to show in seconds whether the load stayed at the right temperature and humidity during transportThe air feels heavy long before anyone measures it: how to make carbon dioxide levels visibleThe freezer showed -18 °C on Monday morning, but what happened over the weekend?The core temperature is what counts, and it only shows if someone has logged itTwo measuring points and the difference between them: how to make the temperature differential in a process visibleFour Measuring Points and a Time-Stamped Log Show Where the Heat in the Process Actually GoesYou Can’t Guess the Core Temperature of a Product: How to Document It Without Standing Next to ItThe fridge looked like it was holding the right temperature, but what happened overnight?

The Sensor Measures Resistance, and So Does the Cable Leading to It

Published August 27, 2026 at 13:42

A platinum sensor does not measure temperature. It is a resistance that changes with temperature, and the instrument converts that resistance into degrees. It’s an excellent principle — but it comes with a built-in complication that dictates how the entire measurement chain must be built: the cable leading to the sensor is also a resistance, and the instrument cannot readily tell the two apart.

TandD TR-8130 handheld Pt100 thermometer with screw terminal

What it is

TandD TR-8130 — a handheld single-channel thermometer for platinum sensors, supplied without a sensor.

  • For Pt100 or Pt1000, in 3- or 4-wire configuration
  • Measuring range -199 to +600 °C, depending on the sensor
  • Screw terminal — the sensor can be swapped in seconds
  • Built-in LCD, battery-powered
  • Compatible with the same Pt sensors as TandD’s logger system

Two ways to get rid of the cable’s resistance

A Pt100 has 100 ohms at zero degrees and changes by just under four tenths of an ohm per degree. A couple of ohms in the cable therefore corresponds to an error of several degrees — and the error is constant, systematic, and looks completely credible.

Pt1000 solves this by making the signal larger. Same platinum, same curve, but ten times the resistance: 1000 ohms at zero degrees and almost four ohms per degree. The cable’s contribution is unchanged, but it now carries ten times less weight. That’s why Pt1000 is the natural choice for long cable runs and simple wiring.

Four-wire connection instead solves it by measuring the cable out of the equation. Two conductors carry the measuring current, two others sense the voltage directly at the sensor — and since virtually no current flows in the sensing leads, no voltage drop occurs in them. The cable’s resistance drops entirely out of the equation, regardless of its length.

Three-wire connection is the compromise: it measures one of the conductors and assumes the other is identical, subtracting the difference. This works well as long as the conductors really are the same length and the same temperature — which they usually are in a factory-made cable, but not always in a spliced installation.

This is the same reasoning behind four-wire measurement of very small resistances: when what you’re measuring is small relative to the path leading to it, the measurement must be separated from the supply.

Why it’s sold without a sensor

The instrument is the reader, the sensor is the measurement. The choice of sensor determines measuring range, response time, durability and form factor — a washable sensor, one with a wide range, or one you already have installed at the facility. The screw terminal lets you swap between them on the spot.

This also makes the instrument useful for something a permanently installed logger can’t do: walking around to check the fixed systems. You bring a known sensor and a handheld reader to the measuring point and compare it against what the facility’s own system reports. If they differ, you know one of them needs attention — and that’s the kind of check that otherwise never gets done.

Three typical use cases

  • Audit rounds: spot checks in food production and pharmaceutical warehouses.
  • Checking fixed systems: comparison against installed sensors and loggers.
  • Temporary measurements: process and energy monitoring where no fixed point exists.

Why it pays off

The platinum sensor’s strength is long-term stability — it drifts very little over the years, which makes it a natural reference to compare other instruments against. But that strength is lost if the measurement chain around it isn’t built for it.

Understanding why Pt1000 exists, and when four wires are needed, is therefore not theory. It’s the difference between a sensor that lives up to its reputation and an installation that throws it away on a couple of ohms of copper.

Read more about the TandD TR-8130 handheld Pt100 thermometer →