Volume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%The air in the fridge doesn’t show the temperature of the vaccineThe Air in the Fridge Lies: How to Monitor the Real Temperature of Your Goods Without False AlarmsHow to show in seconds whether the load stayed at the right temperature and humidity during transportThe air feels heavy long before anyone measures it: how to make carbon dioxide levels visibleThe freezer showed -18 °C on Monday morning, but what happened over the weekend?The core temperature is what counts, and it only shows if someone has logged itTwo measuring points and the difference between them: how to make the temperature differential in a process visibleFour Measuring Points and a Time-Stamped Log Show Where the Heat in the Process Actually GoesYou Can’t Guess the Core Temperature of a Product: How to Document It Without Standing Next to ItThe fridge looked like it was holding the right temperature, but what happened overnight?Volume discount: 5 pcs 2% · 10 pcs 3% · 25 pcs 5%The air in the fridge doesn’t show the temperature of the vaccineThe Air in the Fridge Lies: How to Monitor the Real Temperature of Your Goods Without False AlarmsHow to show in seconds whether the load stayed at the right temperature and humidity during transportThe air feels heavy long before anyone measures it: how to make carbon dioxide levels visibleThe freezer showed -18 °C on Monday morning, but what happened over the weekend?The core temperature is what counts, and it only shows if someone has logged itTwo measuring points and the difference between them: how to make the temperature differential in a process visibleFour Measuring Points and a Time-Stamped Log Show Where the Heat in the Process Actually GoesYou Can’t Guess the Core Temperature of a Product: How to Document It Without Standing Next to ItThe fridge looked like it was holding the right temperature, but what happened overnight?

A pitot tube goes silent when the air moves slowly, a hot-wire doesn’t

Published August 25, 2026 at 09:08

There are two fundamentally different ways to measure how fast air is moving, and the choice between them comes down almost entirely to how slowly it moves.

A pitot tube measures dynamic pressure — the pressure that builds up when the airflow is brought to a stop against the opening of the tube. That pressure grows with the square of the velocity, which means it also shrinks with the square: halve the velocity and the signal drops to a quarter. At just a few metres per second, what remains is so small that it is lost in the instrument’s own noise.

A hot-wire sensor measures something different: how quickly the air cools a heated element. That relationship works well even at very weak air movements, which is why the hot wire takes over exactly where the pitot tube gives up.

SDL350 thermo-anemometer with hot-wire probe and SD logger

What it is

SDL350 — a hot-wire anemometer with a built-in data logger.

  • Telescopic probe with a reach of 215 cm, designed for ventilation ducts and small openings
  • Displays air velocity or airflow and temperature simultaneously on a large LCD
  • Logging to SD card in Excel format — up to 20 million readings, plus 99 manual ones
  • Adjustable logging rate, MIN/MAX, Data Hold, automatic power-off
  • Input for type K and J thermocouples for high temperatures
  • Built-in PC interface, supplied with batteries, SD card and a hard carrying case

Why logging matters more here than on most instruments

Air velocity is rarely a single value — it is a process. The flow in a duct changes when dampers adjust, when other zones open and close, when the fan control responds to a change in pressure and when someone opens a door. A spot measurement therefore describes only the exact moment it was taken, and two people measuring the same diffuser an hour apart can both be right and still disagree.

With logging over 24 hours, the picture changes: you can see whether the diffuser delivers throughout the working day or only between ten and eleven, and you see the pattern rather than the sample. Because the data is saved in Excel format, it can also be attached to a report without any intermediate steps.

The telescopic probe solves a purely geometric problem. The measurement has to be taken inside the duct, at a specific point in the cross-section — and the person measuring has to stand outside. A reach of two metres determines which ducts can be measured at all without opening them up.

And what is worth knowing about the flow value

The instrument can display flow, but flow is calculated from velocity and area. The velocity in a duct is not the same everywhere in the cross-section — friction against the walls makes the air move more slowly near the edges and fastest in the centre. A single measurement point in the middle of the duct therefore gives an average that is too high.

To get a flow value that will stand up to a balancing check, you measure at several points in a grid pattern and calculate the average. The instrument makes this easy by both converting to flow and being able to log each point — but the procedure is something you have to provide yourself. It is not a shortcoming of the instrument, but a property of air in ducts.

Three typical use cases

  • Ventilation balancing: checking diffusers and ducts against design flows.
  • Working environment: draughts and air movement in premises, where velocities are low and the pitot method is not sufficient.
  • Process ventilation: local exhaust points and hoods, where the thermocouple input also covers hot exhaust air.

Why it pays off

Ventilation problems are often troubleshot with arguments instead of measured values: one person feels a draught, another finds the air stuffy, and both experiences are real. Without figures, the corrective action is a guess, and the next guess comes when the complaints return.

A logged measurement series shifts the discussion to what actually happened, hour by hour. And because it is precisely the low velocities — the ones experienced as draughts or as stuffiness — where the pitot method stops working, it is often only with a hot wire that the problem becomes measurable at all.